T1 Mapping · Exercise 3
* The book cites Figure 2.3 and “that blog post”; the WM/GM/CSF curves are Figure 2.2. The simulations on the next slides run live in this page with qmrust, the Rust/WebAssembly port of qMRLab.
From the IR blog notebook (Octave + qMRLab) that generated Figure 2.2’s data. Edit it, then ▶ Run (Ctrl/⌘+Enter): qmrust recomputes it here.
b. Could you use a data analysis pipeline that only has the long-TR IR T1 fitting option available to map T1 in WM, at 1.5T and/or 3T?
| WM T1 (ms) | |
| TR (ms) |
c. Are there any inversion times you should avoid using when designing an IR T1 mapping protocol at 1.5T when using this TR?
Avoid TIs at the null of the tissue you want to measure: magnitude images there are dominated by Rician noise, which can negatively impact the fit under low SNR. For WM at 1.5 T with TR = 3 s, that is TI ≈ 444 ms.
| Part | Answer |
|---|---|
| a | At 1.5 T the WM and GM nulls move earlier: ≈ 451 ms (WM) and 832 ms (GM); CSF is unchanged (2773 ms) |
| b | With TR = 3 s the long-TR fit underestimates WM T1 by ≈ 1.7% at 1.5 T and ≈ 5.8% at 3 T: fine at 1.5 T, biased at 3 T |
| c | Avoid TIs at the WM null, ≈ 444 ms at TR = 3 s: Figure 2.6 shows the STD of the fitted T1 blips up when a TI lands on the null |
Simulations and fits in parts a and b: qmrust (Rust/WebAssembly port of qMRLab), running in the browser.