Data Fitting

Variable Flip Angle

NeuroPoly Lab, Polytechnique Montreal, Quebec, Canada

At first glance, one could be tempted to fit VFA data using a non-linear least squares fitting algorithm such as Levenberg-Marquardt with Eq. 2.5, which typically only has two free fitting variables (T1 and M0). Although this is a valid way of estimating T1 from VFA data, it is rarely done in practice because a simple refactoring of Eq. 2.5 allows T1 values to be estimated with a linear least square fitting algorithm, which substantially reduces the processing time. Without any approximations, Eq. 2.5 can be rearranged into the form y=mx+b\textbf{y}=m\textbf{x}+b Gupta, 1977:

Snsin(θn)=eTRT1Sntan(θn)+C(1eTRT1)\frac{S_n}{ \text{sin}(\theta_n)} = e^{- \frac{TR}{T_1}} \frac{S_n}{ \text{tan}(\theta_n)} + C (1-e^{- \frac{TR}{T_1}})

As the third term does not change between measurements (it is constant for each θn\theta_{n}), it can be grouped into the constant for a simpler representation:

Snsin(θn)=eTRT1Sntan(θn)+C\frac{S_n}{ \text{sin}(\theta_n)} = e^{- \frac{TR}{T_1}} \frac{S_n}{ \text{tan}(\theta_n)} + C

With this rearranged form of Eq. 2.5, T1 can be simply estimated from the slope of a linear regression calculated from Sn/sin(θn)S_{n}/\text{sin}\left( \theta_{n}\right) and Sn/tan(θn)S_{n}/\text{tan}\left( \theta_{n}\right) values:

T1=TRln(slope)T_1 = - \frac{TR}{ \text{ln}(slope)}

If data were acquired using only two flip angles – a very common VFA acquisition protocol – then the slope can be calculated using the elementary slope equation. Figure 2.11 displays both Equations Eq. 2.5 and Eq. 2.8 plotted for a noisy dataset.

Loading...

Figure 2.11:Mean and standard deviation of the VFA signal plotted using the nonlinear form (Eq. 2.5 – blue) and linear form (Eq. 2.8 – red). Monte Carlo simulation details: SNR = 25, N = 1000. VFA simulation details: TR = 25 ms, T1 = 900 ms.

There are two important imaging protocol design considerations that should be taken into account when planning to use VFA: (1) how many and which flip angles to use to acquire VFA data, and (2) correcting inaccurate flip angles due to transmit RF field inhomogeneity. Most VFA experiments use the minimum number of required flip angles (two) to minimize acquisition time. For this case, it has been shown that the flip angle choice resulting in the best precision for VFA T1 estimates for a sample with a single T1 value (i.e. single tissue) are the two flip angles that result in 71% of the maximum possible steady-state signal (i.e. at the Ernst angle) Deoni et al., 2003Schabel & Morrell, 2008.

Time allowing, additional flip angles are often acquired at higher values and in between the two above, because greater signal differences between tissue T1 values are present there (e.g. Figure 2.8). Also, for more than two flip angles, Equations Eq. 2.5 and Eq. 2.8 do not have the same noise weighting for each fitting point, which may bias linear least-square T1 estimates at lower SNRs. Thus, it has been recommended that low SNR data should be fitted with either Eq. 2.5 using non-linear least-squares (slower fitting) or with a weighted linear least-square form of Eq. 2.8 Chang et al., 2008.

Accurate knowledge of the flip angle values is very important to produce accurate T1 maps. Because of how the RF field interacts with matter Sled & Pike, 1998, the excitation RF field (B1+, or B1 for short) of a loaded RF coil results in spatial variations in intensity/amplitude, unless RF shimming is available to counteract this effect (not common at clinical field strengths). For quantitative measurements like VFA which are sensitive to this parameter, the flip angle can be corrected (voxelwise) relative to the nominal value by multiplying it with a scaling factor (B1) from a B1 map that is acquired during the same session:

θcorrected=B1θnominal\theta_{corrected} = B_1 \theta_{nominal}

B1 in this context is normalized, meaning that it is unitless and has a value of 1 in voxels where the RF field has the expected amplitude (i.e. where the nominal flip angle is the actual flip angle). Figure 2.12 displays fitted VFA T1 values from a Monte Carlo dataset simulated using biased flip angle values, and fitted without/with B1 correction.

Figure 2.12:Mean and standard deviations of fitted VFA T1 values for a set of Monte Carlo simulations (SNR = 100, N = 1000), simulated using a wide range of biased flip angles and fitted without (blue) or with (red) B1 correction. Simulation parameters: TR = 25 ms, T1 = 900 ms, θnominal\theta_{nominal} = 6° and 32° (optimized values for this TR/T1 combination). Notice how even after B1 correction, fitted T1 values at B1 values far from the nominal case (B1 = 1) exhibit larger variance, as the actual flip angles of the simulated signal deviate from the optimal values for this TR/T1 (Deoni et al. 2003).

Figure 2.13 displays an example VFA dataset and a B1 map in a healthy brain, along with the T1 map estimated using a linear fit (Equations Eq. 2.8 and Eq. 2.9).

Loading...

Figure 2.13:Example variable flip angle dataset and B1 map of a healthy adult brain (left). The relevant VFA protocol parameters used were: TR = 15 ms, θnominal\theta_{nominal} = 3° and 20°. The T1 map (right) was fitted using a linear regression (Equations Eq. 2.8 and Eq. 2.9).

References
  1. Gupta, R. K. (1977). A new look at the method of variable nutation angle for the measurement of spin-lattice relaxation times using fourier transform NMR. Journal of Magnetic Resonance (1969), 25(1), 231–235. 10.1016/0022-2364(77)90138-x
  2. Deoni, S. C. L., Rutt, B. K., & Peters, T. M. (2003). Rapid combined T1 and T2 mapping using gradient recalled acquisition in the steady state. Magnetic Resonance in Medicine, 49(3), 515–526. 10.1002/mrm.10407
  3. Schabel, M. C., & Morrell, G. R. (2008). Uncertainty in T1 mapping using the variable flip angle method with two flip angles. Physics in Medicine and Biology, 54(1), N1–N8. 10.1088/0031-9155/54/1/n01
  4. Chang, L.-C., Koay, C. G., Basser, P. J., & Pierpaoli, C. (2008). Linear least-squares method for unbiased estimation of T1 from SPGR signals. Magnetic Resonance in Medicine, 60(2), 496–501. 10.1002/mrm.21669
  5. Sled, J. G., & Pike, G. B. (1998). Standing-wave and RF penetration artifacts caused by elliptic geometry: an electrodynamic analysis of MRI. IEEE Transactions on Medical Imaging, 17(4), 653–662. 10.1109/42.730409