qMRLab logoQuantitative MRI mOOC

B0 Mapping · Exercise 5-3

Shimming

In this exercise, we will cover:

  1. How to find the shim currents that make the field more uniform
  2. How your shim compares with the optimal one
  3. Volume-wise and slice-wise shimming

Exercise 5-3 · (a)

Shimming the B0 map

Find the shim currents that optimize the shim, that is, that minimize the mean and the SD of B0.

Input

B0 before shimming

Output

B0 after shimming
Mean
SD

Exercise 5-3 · (b)

Your shim and the optimal shim

Submit your shim on the previous slide.

Your shim Optimal shim

Your shim
Optimal shim
Yours − optimal

Exercise 5-3 · (c)

Volume-wise vs slice-wise shimming

Find the shim currents that minimize the mean and the SD of B0: first over the whole brain, then over axial slice 35 only.

Input

B0 as acquired
Shim over

Output

B0 after shimming

Mean
SD

Data: Chan K-S, Marques JP, Spincemaille P, Wang Y. QSM consensus example data (Siemens). Zenodo, doi:10.5281/zenodo.7795492, CC BY 4.0; echoes 1 and 2, 2 mm, phase difference unwrapped in 3D.

Exercise 5-3 · (d)

Volume-wise and slice-wise shims

Submit your shim on the previous slide.

Yours, volume

Optimal, volume

Yours, slice 35

Optimal, slice 35

Volume − slice 35

Exercise 5-3 · Summary

Summary

  1. B0 maps let us compute the shim currents that reduce the field inhomogeneity.
  2. Volume-wise shimming minimizes the mean and the SD of Δf over the whole masked volume; slice-wise shimming does it over a single slice only.
  3. Slice-wise shimming can give better shims for multislice acquisitions, with the shims updated for each slice.
Shimming Toolbox

To compute shims for your own data, you can use Shimming Toolbox, an open-source tool developed at NeuroPoly.

NeuroPoly