B0 Mapping · Exercise 5-1

Dual-echo B0 mapping: a single-voxel example

In this exercise, we will cover:

  1. How to compute B0 for a voxel of pure water
  2. How the fat signal affects the B0 measurement
  3. How to mitigate the effects of fat on the B0 measurement

Exercise 5-1Double-echo B0 mapping in a single voxel

TE:ms
Freq. offset:Hz
B0:T
$$\Delta B_0\,[\mathrm{T}] = \frac{\phi(TE_2) - \phi(TE_1)}{\gamma\,\Delta TE}$$
Lecture slide 10: dual echo ↗
φ(TE1)rad φ(TE2)rad ΔTEs

Question 1a

You are scanning on a 3 T scanner, and your sequence has a minimum TE of 2 ms. Compute the difference in phase between that measurement and a measurement at 4 ms, for two cases:

1.a freq. offset of 0 ppm:Δφ =° 2.a freq. offset of 0.5 ppm:Δφ =°

Question 1b

What is the measured ΔB0 for each case, in µT?

1.a freq. offset of 0 ppm:ΔB0 =µT 2.a freq. offset of 0.5 ppm:ΔB0 =µT

Exercise 5-1 · Summary

Summary

  1. Two echoes give ΔB0 from the change in phase between them: ΔB0 = Δφ / (γ ΔTE), with Δφ in radians and ΔTE in seconds.
  2. Fat precesses 3.4 ppm slower than water (−434 Hz at 3 T), so in a voxel with fat the measured phase is that of the sum of the two, and the B0 measured is biased.
  3. Echoes taken when water and fat are in phase (every 2.30 ms at 3 T) remove that bias.